Terms of the MRI examination

  • Has the SureScan simulation system been implanted at least six weeks ago?
  • Has the SureScan device been implanted in the chest?
  • The effective stimulation threshold is at < 0.2 V at a pulse duration of 0.4 ms.
  • The electrode impedance is in the normal measurement range (over 1'500 Ω or below 200 Ω).
  • No other devices, discontinued or additional electrodes, adapters or extensions.

For patients in possession of a pass for the MRI Conditional SureScan System the implanting cardiologist confirmed by issueing the pacemaker pass that no other (active or discontinued) medical devices, electrodes, electrode extensions or electrode adapters are implanted.

Are all requirements fullfilled the pacemaker has to be programmed into SureScan-mode for the MRI examination.

Further details are found in the manual.
Information about MRI examinations in combination with the Advisa DR MRI™ SureScan® A3DR01 and the limited MRI-secure SureScan® electrodes.

Patient is ready and may be referred to the radiology

After the MRI examination

  • Program the MRI SureScan function to off.

Further details are found in the manual.
Information about MRI examinations in combination with the Advisa DR MRI™ SureScan® A3DR01 and the limited MRI-secure SureScan® electrodes.

Pacemaker

A3DR01 Advisa DR MRI SureScan UNI/BI/IS-1 PZK
A3SR01 Advisa SR MRI SureScan UNI/BI/IS-1 PTR
EMDR01 EnRythm DR MRI Surescan UNI/BI/IS-1 PTA
EN1DR01 EnSura DR MRI UNI/BI/IS-1 PZW
EN1SR01 Ensura SR MRI UNI/BI/IS-1 PTZ

Manual about pacemakers

Atrium sonde

5086MRI CAPSUREFIX MRI™ IS-1/BI LFP
Lengths: 45, 52, 58 cm
5076 CapSureFix Novus MRI™ IS-1/BI PJN
Lengths: 35, 45, 52, 58, 65, 85 cm
4574 CAPSURE SENSE MRI™ IS-1/BBI BBE
Lengths: 45, 53 cm
4074 CAPSURE SENSE MRI™ IS-1/BI BBD
Lengths: 52, 58 cm

Ventricle sonde

4074 CAPSURE SENSE MRI™ IS-1/BBI BBD
Lengths: 52, 58 cm
5086MRI CAPSUREFIX MRI™ IS-1/BI LFP
Lengths: 45, 52, 58 cm
5076 CapSureFix Novus MRI™ IS-1/BI PJN
Lengths: 35, 45, 52, 58, 65, 85 cm